Device Overview

The ICU Medical Plum 360 infusion pump delivers precise, programmable rates, integrating drug concentration calculations for accurate dosing. Its user‑friendly interface, alarm system, and MedNet compatibility streamline workflow while enhancing patient safety. Compact design suits ICU settings. 2026.

Setup and Installation

Before first use, verify the Plum 360 is powered by a 24‑V DC source or the supplied AC adapter. Connect the power cable to the rear connector, ensuring the polarity indicator aligns with the device. Attach the infusion set to the patient port, checking for a secure seal and no kinks. Verify the tubing is free of air bubbles and that the IV bag is properly positioned. The pump’s integrated drug library requires initialization: press the Menu button, navigate to Library, and select Load. If the device is MedNet‑ready, connect the RJ‑45 cable to the MedNet port and configure network settings via the Network menu. Perform a self‑check by selecting Diagnostics → Run; the display will show a green checkmark when all systems are functional. Confirm the device clock is set correctly by selecting Clock and entering the current time. Set the infusion parameters: choose Rate, input volume or concentration, and confirm with Enter. The pump will calculate the appropriate flow rate and display it in mL/hr. Review the infusion summary on the screen and verify all values match the prescription. Press the Start button to begin the infusion. Monitor the status screen for any alarms or errors. If an alarm appears, consult the alarm code list and take corrective action. For routine maintenance, refer to the cleaning section. Ensure the device is stored in a clean, dry area with temperature between 15°C and 30°C. Label the device with the installation date and technician initials. Keep a log of all installation steps for audit purposes. The device is now ready for clinical use.

Operation Modes

The Plum 360 offers three primary infusion modes: Rate‑Based, Volume‑Based, and Concentration‑Based. In Rate‑Based mode, the user enters a desired flow rate in mL/hr; the pump calculates the necessary pressure to maintain that rate. Volume‑Based mode requires the total volume to be delivered and the time window; the device then auto‑adjusts the rate to meet the target. Concentration‑Based mode allows the clinician to specify drug concentration and desired infusion volume; the pump automatically determines the correct flow rate and alerts if the concentration exceeds safe limits. Each mode supports bolus and continuous delivery, and the user can toggle between Manual and Auto adjustment of pressure. The device’s Smart‑Ramp feature gradually increases or decreases flow to avoid sudden changes. For advanced protocols, the Multi‑Step mode lets the user program a sequence of rates or volumes, each with its own duration. The pump’s Alarm‑Triggered mode pauses infusion when a critical alarm occurs and resumes automatically once the issue is resolved. All modes are accessible via the Mode menu, and the current mode is displayed on the main screen. The system logs each mode change for audit purposes. The user can also set Safety Limits for maximum and minimum flow rates, and the pump will enforce these limits in all modes. The firmware supports OTA updates, enabling clinicians to deploy patches and features without reprogramming, thus maintaining compliance with evolving daily regulations.

Alarm and Troubleshooting

The Plum 360’s alarm system is designed to provide immediate, actionable notifications for a wide range of operational events. The device displays a concise alarm code, a brief description, and a recommended corrective action. For example, alarm code N185 indicates a proximal occlusion; the user should check the tubing for kinks or blockages and clear the line. Table 6‑1 in the Technical Service Manual lists all operational alarm messages, their meanings, and suggested actions. Common alarms include N186 (distal occlusion), N187 (air in line), and N188 (pump failure). Each alarm triggers a visual LED indicator and an audible tone. The user can acknowledge an alarm by pressing the ACK button; the alarm will remain active until the underlying issue is resolved or the user presses ACK again. If an alarm persists after corrective action, the pump will automatically enter a safe mode, pausing infusion until the problem is resolved. Troubleshooting steps follow a systematic approach: 1) Verify the alarm code and description; 2) Inspect the infusion set, tubing, and drug bag; 3) Check for air, kinks, or occlusions; 4) Ensure the pump’s power supply is stable; 5) Review the infusion program settings. If the issue remains unresolved, consult the detailed troubleshooting guide in the manual or contact technical support. The pump’s firmware logs all alarms and corrective actions, allowing for post‑event analysis and quality improvement. Maintaining a clear, step‑by‑step troubleshooting protocol helps minimize downtime and ensures patient safety. All alarm logs are retained for 30 days and can be exported via the MedNet interface for audit purposes. If unresolved, the pump enters safe mode!!

Maintenance and Safety

The ICU Medical Plum 360 requires maintenance to preserve accuracy and reliability. checks include verifying the power supply, inspecting the infusion set for kinks or air, and confirming the drug bag is secured. The device’s firmware logs all maintenance actions; these logs should be reviewed monthly to detect trends.

Cleaning and sterilization are performed each use. Remove the infusion set, rinse the tubing with sterile water, and wipe the pump’s touch screen with a 70% isopropyl solution. The pump’s internal components are not exposed to water, so only external surfaces are cleaned. After cleaning, allow the device to air‑dry for at least 5 minutes before the next use.

Service intervals are specified in the Technical Service Manual. pump’s internal calibration should be performed every 12 months or after 500 infusion hours, whichever occurs first. Replace battery every 24 months if the pump is not in continuous use. Filter cartridge must be replaced every 6 months or after 300 infusion hours.

Safety precautions include keeping the pump away from extreme temperatures (0–40 °C), avoiding exposure to magnetic fields, ensuring the device is on a stable, flat surface. Never use the pump if the display indicates a fault or if the alarm remains active after corrective action. Always follow the manufacturer’s instructions for drug compatibility and infusion set selection.

In case of a malfunction, document the event, perform the recommended corrective action, contact ICU Medical technical support. Pump’s safety features—such as automatic occlusion detection and pressure monitoring—are designed to prevent medication errors and protect patient well‑being Record maintenance actions!

Software and Connectivity

The ICU Medical Plum 360 supports MedNet integration, allowing real‑time data transfer via Ethernet or USB. Firmware updates are delivered through the MedNet portal or USB drive, ensuring compliance with standards. The pump’s touchscreen interface displays infusion parameters and logs for audit trails.

Physical Inspection and Assembly

The Plum 360 features a safety system that includes a check valve and an occlusion sensor. Verify the check valve and occlusion sensor are unobstructed. The pump’s display will show checkmarkonce the assembly is complete. Clean the exterior with cloth and detergent; avoid abrasive cleaners. After cleaning, allow the device to dry for 10 minutes before powering on. Keep record of all maintenance activities in the maintenance log. Follow the manufacturer’s recommended calibration interval of 90 days. Ensure that the device’s battery backup is charged and that the emergency power switch is functional. Verify that the device’s alarm volume is set to level audible to staff in the ICU environment. Once all steps are completed, the Plum 360 is ready for clinical use.

Basic Rate Infusion

The ICU Medical Plum 360’s Basic Rate Infusion mode allows clinicians to set a precise infusion rate by entering either a desired volume per hour or a drug concentration and total volume. To initiate a basic infusion, press the Start button, then use the numeric keypad to input the target rate in mL/hr. The display will confirm the rate and calculate the total volume remaining. If a drug concentration is required, select the Concentration option, enter the drug concentration in mg/mL, and the pump will automatically compute the infusion rate needed to deliver the prescribed dose over the specified time. The pump’s built‑in safety features monitor for occlusion, air, and low‑volume conditions; any deviation triggers an audible alarm and a visual warning on the screen. The user can pause, resume, or stop the infusion at any time by pressing the Pause or Stop buttons. All infusion events are logged in the internal memory, which can be exported via the USB port for audit purposes. The device supports a maximum infusion rate of 10,000 mL/hr and a minimum of 0.1 mL/hr, ensuring flexibility for both large‑volume fluid therapy and precise drug delivery. The interface also displays the remaining time and volume, allowing staff to anticipate when the infusion will complete. For multi‑step infusions, the user can program a sequence of rates, each with its own start time and duration, using the Sequence menu. The pump’s automatic rate calculation reduces dosing errors MedNet for remote now monitoring.

Advanced Rate Calculation

The ICU Medical Plum 360 Advanced Rate Calculation mode empowers clinicians to program complex infusion regimens that adjust automatically to changing patient parameters. By selecting the Advanced option, the user can input a target drug concentration, total volume, and desired infusion duration. The pump’s algorithm calculates the precise hourly rate required to achieve the target concentration, taking into account the drug’s solubility and the infusion bag’s volume. Once the parameters are entered, the display shows the calculated rate, the remaining volume, and the projected completion time. The device supports stepwise rate changes, allowing a clinician to schedule multiple phases—such as an initial loading dose followed by a maintenance infusion—within a single program; Each phase can be defined by its own start time, duration, and rate, and the pump will transition seamlessly between them. The Advanced mode also offers a “Target Concentration” feature, which continuously monitors the drug concentration in the infusion line and adjusts the flow rate to maintain the prescribed concentration, compensating for any changes in bag volume or temperature. This feature is particularly useful for delivering chemotherapy agents or other concentration‑sensitive medications. The pump logs all infusion parameters and events in its internal memory, enabling post‑procedure review and audit. Data can be exported via the USB port or transmitted over the MedNet network to a central monitoring

Alarm Codes and Corrective Actions

The ICU Medical Plum 360 features a robust alarm system that notifies users of critical events. Each alarm is identified by a unique code and a brief description, followed by a recommended corrective action. The table below lists the most frequently encountered alarms.

  • N185 – Proximal Occlusion: Proximal blockage detected.
    Action: Inspect and clear proximal tubing; replace if necessary.
  • N186 – Distal Occlusion: Distal blockage detected.
    Action: Check distal tubing for kinks; replace if needed.
  • N187 – Air in Line: Air detected.
    Action: Flush line with saline until bubbles disappear.
  • N188 – Low Reservoir: Reservoir volume low.
    Action: Replace bag with a full reservoir.
  • N189 – Pump Failure: Motor failure.
    Action: Power‑cycle pump; if alarm persists, contact service.
  • N190 – Battery Low: Battery voltage low.
    Action: Replace battery with a fresh unit.

All alarms are displayed on the main screen and emit an audible tone. Users acknowledge alarms by pressing the ACK button, which clears the message but does not resolve the underlying issue. Persistent alarms must be investigated promptly to prevent infusion interruptions or medication errors. The pump logs each alarm event with a timestamp and corrective action for compliance purposes.

Refer to manual for troubleshooting.

Cleaning and Sterilization Procedures

The ICU Medical Plum 360 infusion pump requires routine cleaning and sterilization to maintain sterility and performance. Follow these steps before each use:

  • Disassemble all removable components: tubing, reservoir, and needleless connectors. Separate the pump head from the base.
  • Wipe the external surfaces with a 70% isopropyl alcohol wipe. Use a lint‑free cloth to avoid residue.
  • Clean the reservoir and tubing with a sterile, non‑ionic detergent solution. Rinse thoroughly with sterile water to remove detergent.
  • Inspect all parts for cracks, wear, or contamination. Replace any damaged component.
  • Dry all parts in a clean, dry environment. Do not use a heat source that exceeds 50 °C.

For sterilization of the tubing and reservoir, use a validated steam‑sterilization cycle at 121 °C for 15 minutes. Ensure the pump is fully powered off and disconnected from any power source before sterilization. Do not immerse the pump base in any liquid. After sterilization, allow components to cool to room temperature before reassembly. Store the pump in a clean, dry cabinet, protected from dust and direct sunlight. Perform a visual inspection before each use to confirm sterility and integrity.

All cleaning and sterilization procedures must be documented in the device log. Record the date, time, personnel, and any deviations from the standard protocol. This documentation supports assurance and compliance.

MedNet Device Compatibility

The ICU Medical Plum 360 is fully integrated with MedNet, enabling seamless communication with compatible monitoring and electronic medical record systems. The pump supports MedNet 2.0 and 3.0 protocols, allowing bidirectional data exchange for infusion parameters, alarms, and device status. Users can configure the pump via the MedNet interface to receive real‑time updates on infusion rates, volume delivered, and battery status. Alarm codes are transmitted to the central monitor, ensuring rapid response by clinical staff. The device also supports secure data logging, with encrypted transmission to the MedNet server for audit trails. Compatibility extends to all ICU Medical MedNet‑approved accessories, including the MedNet‑enabled infusion set, reservoir, and syringe adapter. Firmware updates are delivered through MedNet, guaranteeing that the pump remains current with the latest safety features and regulatory changes. Integration with MedNet enhances workflow efficiency, reduces medication errors, and improves patient safety across the ICU environment. The MedNet integration also provides automated data backup to the hospital’s central server enabling real‑time analytics and trend monitoring for critical care teams. Additionally, the pump’s firmware supports secure over‑the‑air updates, ensuring that all devices remain compliant with the latest regulatory standards without requiring manual intervention. Compliance logs are automatically archived, providing audit trails that meet 13485 requirements.

Service and Calibration Intervals

According to the ICU Medical Plum 360 technical service manual, routine service and calibration are essential for maintaining accurate infusion delivery and ensuring patient safety. The pump requires a full functional check and calibration of the flow sensor every 12 months or after any major component replacement. The infusion rate accuracy must be verified quarterly using a calibrated flow meter and a standard 10 mL syringe. Service includes inspection of the power supply, battery, and all electrical connectors for corrosion or loose contacts, and a leak test on the infusion set. The manual specifies that the flow sensor should be calibrated against a reference standard with a tolerance of ±0.5 %. The calibration process involves setting the pump to a known flow rate, recording the sensor output, and adjusting the calibration coefficient until the error falls within the acceptable range. All calibration data must be logged in the device’s internal memory and backed up to the MedNet server for audit purposes. Calibration should be documented in the patient chart and cross‑checked with pharmacy records. The technician should verify that alarm thresholds for occlusion, low battery, and low fluid volume are correctly programmed and that audible and visual alarm outputs function properly. Finally, the manual recommends cleaning the pump’s external surfaces with a mild, non‑abrasive detergent and a dry microfiber cloth, followed by a sterility check of the infusion set.